Street Address: 9/ 10 Jintai Building Nanping Street Nanan District
City: Chongqing
Province: Chongqing
Country/Region: China
GDYJ-501 Transformer Oil Breakdown Voltage Tester is developed according to the national standard GB-86 Insulating Oil Dielectric Strength Testing Method .
1.1 Introduction » GDYJ-501 Transformer Oil Breakdown Voltage Tester is developed according to the national standard GB-86 Insulating Oil Dielectric Strength Testing Method . LCD screen can display test procedure and test result. All the human interaction operations complete by the rotating mouse . » It is very easy for customers to measure the breakdown voltage of all kinds of oil by GDYJ-501 BDV Oil Breakdown Voltage Tester. You can print the test result by the micro printer internal the tester. Of course, you can upload the test data to PC from RS232 for data analyzing. The special circuit design makes the tester has a good performance at stability and anti-noise. 1.2 Functions of oil BDV tester 1 Breakdown test of user-defined 2 Proof test 3 Calibration interface for customers 4 Unloading data to PC and making word report automatically. 5 Result print 1.3 Technique Index 1 Output voltage AC 0~80KV/AC 0~100KV 2 Rate of voltage rise 2KV/s±10% 3 Measuring precision 3%RDG+0.3%FS 4 Test times setting 1~9 5 Mixing time setting 0~99 second 6 Waiting time setting 0-9 minute 59second 7 Maximum storage capacity 100 groups of test data 8 Test standards IEC60156 9 Display 256*128 LCD 10 Rated power supply AC 220V±10% 50Hz/60Hz 11 Environment temperatures 0-50°C 12 Relative humidity ≤80%RH 13 Size 420×380×410mm3 14 Weight 28kg