Scope
Compliance with SH / T 0251 "Determination of base number of petroleum products (perchloric acid potentiometric titration)" and GB / T 7304 "Petroleum products and lubricants and values assay (potentiometric titration)" standard laboratory requirements.
Features
1, the use of liquid crystal display, English operation interface, can display information about the test parameters and measurement results
2, with pre-titration, preset end point titration, blank titration or manual titration functions, and can generate special titration mode according to user habits
3, the choice of different electrodes can be: acid-base titration, redox titration, precipitation titration, complexation titration, and other non-aqueous titration and pH measurement
4, stirring system uses PWM modulation technology, software, speed, low noise
5, perchloric acid titration system using anti-corrosion materials, can be a variety titration
Technical Parameters
First, the instrument level: 0.05
1, measuring range:
pH: (0.00 ~ 14.00) pH
mV: (- 1800 ~ 1800.0) mV
Temperature: (- 5.0 ~ 105.0) ° C
2, Resolution:
pH: 0.01pH
mV: 0.1mV
Temperature: 0.1 ° C
3, the basic error:
pH: ± 0.01pH ± 1 word
mV: ± 0.03% FS
Temperature: ± 0.3 ° C ± 1 word
4, Burette capacity Tolerance:
10ml Buret: ± 0.025ml
20ml Buret: ± 0.035ml
5, Burette resolution:
10ml Buret: 1/10000
20ml burette: 2/10000
6, burette infusion or infusion rate: (50 ± 10) s (Buret full degrees)
7, titration analysis Repeatability: 0.2%
Titration control sensitivity: ± 2 mV
8 Stability: (± 0.3mV ± 1 word) / 3h
9, power source: AC (220 ± 22) V, (50 ± 1) Hz
10, Dimensions (mm): 340 × 400 × 400
10. Instrument Weight: 10 kg
Related Product Tags:
turbocharger balancing machine , balancing machine , battery tester , bench , body bench , body repair bench , repair bench , testing bench , auto bench , brake tester , Clutch Tester , common rail system , common rail test stand , common rail test bench , common rail tester , diesel fuel pump , diesel test bench , four wheel aligner